The accuracy of two new 4-h identification systems for anaerobes, the AN-IDENT (Analytab Products, Plainview, N.Y.) and the RapID ANA (Innovative Diagnostic Systems, Inc., Atlanta, Ga.) was compared with that of the API 20A system (Analytab Products). A total of 132 clinical isolates were tested in each of the three systems. The overall accuracies at the genus and species level for the three systems were: API 20A, 68.9 and 56.8%, respectively; AN-IDENT, 90.2 and 73.5%; and RapID ANA, 93.9 and 81.8%. Improved identification of anaerobes with the AN-IDENT and the RapID ANA systems was observed for isolates of the genus Fusobacterium, Clostridium species other than Clostridium perfringens, non-spore-forming bacilli, and isolates of the genus Peptostreptococcus. Reproducibility studies demonstrated that the results of the individual test reactions in all three identification systems were reproducible when the interpretive guidelines of the manufacturer were followed precisely.

译文

将两种新的4小时厌氧菌识别系统的准确性,即AN-IDENT (Analytab Products,纽约州Plainview) 和RapID ANA (Innovative Diagnostic systems,inc.,乔治亚州亚特兰大) 与API 20A系统 (Analytab产品)。在这三个系统中的每个系统中总共测试了132个临床分离株。这三个系统在属和种级别的总体精度分别为: API 20A,68.9和56.8%; A-IDENT,90.2和73.5%; 和快速ANA,93.9和81.8%。对于梭菌属,除产气荚膜梭状芽孢杆菌以外的梭状芽孢杆菌属,非孢子形成杆菌和消化链球菌属的分离株,观察到an-ident和快速ANA系统对厌氧菌的鉴定得到了改善。可重复性研究表明,当精确遵循制造商的解释准则时,所有三个识别系统中单个测试反应的结果都是可重现的。