We sporadically experienced three paediatric cases of atypical upswing at the initial part of phase III in the capnograms via side-stream capnometer immediately following endotracheal intubation and mechanical ventilation. No fault was found in the monitor or anaesthetic system including breathing circuits, carbon dioxide (CO2 ) sampling tube, water trap, and unidirectional valves. The upsurge of CO2 disappeared with increasing the respiratory rate; however, it reappeared with decreasing the respiratory rate and vice versa. We experimentally reproduced the phenomenon of overshooting CO2 measurement after the luer lock connection of the sample gas line at the water trap had been unscrewed a little bit.

译文

我们在气管插管和机械通气后立即通过侧流二氧化碳计在二氧化碳图的III期初期偶尔经历了3例非典型上升的儿科病例。在监护仪或麻醉系统 (包括呼吸回路,二氧化碳 (CO2) 采样管,集水器和单向阀) 中未发现故障。二氧化碳的激增随着呼吸频率的增加而消失; 然而,它随着呼吸频率的降低而再次出现,反之亦然。我们通过实验再现了在疏水阀上的样品气体管线的luer锁连接稍微拧开后的CO2测量超调现象。