The aim of this study is to develop a postlaryngectomy airway climate explorer (ACE) for assessment of intratracheal temperature and humidity and of influence of heat and moisture exchangers (HMEs). Engineering goals were within-device condensation prevention and fast response time characteristics. The ACE consists of a small diameter, heated air-sampling catheter connected to a heated sensor house, containing a humidity sensor. Air is sucked through the catheter by a controlled-flow pump. Validation was performed in a climate chamber using a calibrated reference sensor and in a two-flow system. Additionally, the analyser was tested in vivo. Over the clinically relevant range of humidity values (5-42 mg H2O/l air) the sensor output highly correlates with the reference sensor readings (R2 > 0.99). The 1-1/e response times are all <0.5 s. A first in vivo pilot measurement was successful. The newly developed, verified, fast-responding ACE is suitable for postlaryngectomy airway climate assessment.

译文

:本研究的目的是开发一种喉后切除术气道气候探测器(ACE),以评估气管内温度和湿度以及热湿交换器(HME)的影响。工程目标是防止设备内凝结和具有快速响应时间的特性。 ACE由一个小直径的加热空气采样导管组成,该导管连接到包含湿度传感器的加热传感器外壳。空气由可控流量泵通过导管吸入。验证是在气候室中使用已校准的参考传感器和双流系统进行的。另外,分析仪在体内进行了测试。在临床上相关的湿度值范围(5-42 mg H2O / l空气)中,传感器输出与参考传感器读数高度相关(R2> 0.99)。 1-1 / e响应时间均<0.5 s。首次体内试点测量成功。新开发的,经过验证的,快速响应的ACE适用于喉切除术后气道气候评估。