BACKGROUND:New 2007 European Society of Hypertension guidelines recommend measuring arterial stiffness in patients with arterial hypertension, suggesting a carotid-femoral pulse wave velocity over 12 m/s as an estimate of subclinical organ damage. Considering this cutoff point, it is worth exploring whether or not there are significant differences in results obtained using various techniques for measuring aortic pulse wave velocity. The aim of the study was to compare aortic pulse wave velocity measurements using Complior, SphygmoCor, and Arteriograph devices, and to assess the effect of pulse wave transit time and traveled distance on pulse wave velocity values. METHODS:Aortic pulse wave velocity was measured on a single visit, using these devices, in randomized order, in a group of 64 patients with grade 1 or 2 arterial hypertension. RESULTS:Aortic pulse wave velocity measured using Complior (10.1 +/- 1.7 m/s) was significantly higher than that obtained using SphygmoCor (8.1 +/- 1.1 m/s) or Arteriograph (8.6 +/- 1.3 m/s). No differences were noted between pulse wave velocity measurements using SphygmoCor and Arteriograph. Between-method comparison revealed that differences in traveled distance were significant: Complior versus Arteriograph [0.09 m, Confidence interval (CI): 0.08-0.12 m, P < 0.05], Complior versus SphygmoCor (0.15 m, CI: 0.13-0.16 m, P < 0.05), Arteriograph versus SphygmoCor (0.05 m, CI: 0.03-0.07 m, P < 0.05). No between-method differences were found for transit times. CONCLUSION:Differences in pulse wave velocity obtained by compared devices resulted primarily from using various methods for measuring traveled distance. It appears reasonable to establish uniform principles for the measurement of traveled distance. Because a large number of prognosis/survival studies used direct distance between carotid and femoral sites of pulse wave recording, this distance should be mostly recommended.

译文

摘要背景:2007年新的欧洲高血压学会指南建议测量高血压患者的动脉僵硬度,建议将颈股脉搏波速度超过12 m / s作为亚临床器官损伤的估计值。考虑到这个截止点,值得探讨使用各种测量主动脉脉搏波速度的技术所获得的结果是否存在显着差异。该研究的目的是比较使用Complior,SphygmoCor和Arteriograph装置测量的主动脉脉搏波速度,并评估脉搏波传播时间和行进距离对脉搏波速度值的影响。
方法:随机选择64例1级或2级动脉高压患者使用这些设备,随机访问一次主动脉脉搏波速度。
结果:使用Complior(10.1 /-1.7 m / s)测量的主动脉脉搏波速度显着高于使用SphygmoCor(8.1 /-1.1 m / s)或动脉造影仪(8.6 /-1.3 m / s)测得的主动脉脉搏波速度。使用SphygmoCor和Arteriograph测量脉搏波速度之间没有发现差异。方法之间的比较显示,行进距离差异很大:Complior与动脉造影仪[0.09 m,置信区间(CI):0.08-0.12 m,P <0.05],Complior与SphygmoCor(0.15 m,CI:0.13-0.16 m, P <0.05),动脉造影仪与SphygmoCor的比较(0.05 m,CI:0.03-0.07 m,P <0.05)。在运输时间上未发现方法间差异。
结论:比较装置获得的脉搏波速度差异主要是由于使用各种测量行进距离的方法引起的。建立行进距离测量的统一原理似乎是合理的。由于大量的预后/生存研究均使用脉搏波记录的颈动脉和股骨部位之间的直接距离,因此应主要建议使用该距离。

+1
+2
100研值 100研值 ¥99课程
检索文献一次
下载文献一次

去下载>

成功解锁2个技能,为你点赞

《SCI写作十大必备语法》
解决你的SCI语法难题!

技能熟练度+1

视频课《玩转文献检索》
让你成为检索达人!

恭喜完成新手挑战

手机微信扫一扫,添加好友领取

免费领《Endnote文献管理工具+教程》

微信扫码, 免费领取

手机登录

获取验证码
登录