Discrepancies of one pH unit and more have been observed after a few days, between continuous on-line in situ pH measurements and instant off-line pH measurements during anaerobic digestion of an agroindustrial wastewater. Concomitantly, the electrical resistance across the porous diaphragm of the on-line electrode increased, and a black clogging developed on its diaphragm. Measurements of the relative liquid junction potential in KCl or Na(2)S solutions excluded that high concentrations of ions such as, K(+), Na(+), Cl(-), HS(-), or S(2-) were the major cause of the drifts in pH values. It has been possible to limit the rapid increase of the liquid junction potential and the black clogging formation on the porous diaphragm either by acidification and/or by overpressurization of the electrode-filling liquid. Continuous on-line in situ pH values consistent with instant off-line pH values over long periods of time have been obtained with a newly designed pH sensor in which a special jellied electrode filling replaced the porous diaphragm.

译文

几天后,在农业工业废水的厌氧消化过程中,连续在线原位pH测量与即时离线pH测量之间观察到一个pH单位或更多pH单位的差异。同时,在线电极的多孔隔膜上的电阻增加,并且隔膜上出现了黑色堵塞。KCl或Na(2)S溶液中相对液体结电位的测量排除了高浓度的离子,如,K(+),Na(+),Cl(-),HS(-),or S(2-) 是ph值漂移的主要原因。通过酸化和/或电极填充液体的过压,可以限制液体结电位的快速增加和多孔隔膜上的黑色堵塞形成。使用新设计的pH传感器,可以获得连续的在线原位ph值,与长时间的离线ph值一致,其中使用特殊的凝胶状电极填充代替了多孔隔膜。